Title of article :
An enhanced method to calibrate vibrating tube densimeters
Author/Authors :
Bouchot، نويسنده , , C. and Richon، نويسنده , , D.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
20
From page :
189
To page :
208
Abstract :
The conversion of vibrating periods to densities through the calibration of the Anton Paar DMA 512 vibrating tube densimeter (VTD) to perform P–ρ–T measurements on fluids is examined. After reviewing various calibration models with their detailed internal characteristics, a new one is proposed that accurately extends the use of the densimeter over a wide range of densities, including vapors under low pressures, compressed liquids and near critical fluids. A good accuracy is obtained by fitting only two mechanically defined parameters on liquid reference density data. Detailed considerations on the mechanical model and on the calibration procedure are exposed and illustrated through sample results.
Keywords :
Vibrating tube , densimeter , Calibration , P–?–T
Journal title :
Fluid Phase Equilibria
Serial Year :
2001
Journal title :
Fluid Phase Equilibria
Record number :
2033828