Title of article
On the tip calibration for accurate modulus measurement by contact resonance atomic force microscopy
Author/Authors
Passeri، نويسنده , , D. and Rossi، نويسنده , , M. and Vlassak، نويسنده , , J.J.، نويسنده ,
Pages
10
From page
32
To page
41
Abstract
Accurate quantitative elastic modulus measurements using contact resonance atomic force microscopy require the calibration of geometrical and mechanical properties of the tip as well as the choice of a suitable model for describing the cantilever-tip-sample system. In this work, we demonstrate with both simulations and experiments that the choice of the model influences the results of the calibration. Neglecting lateral force results in the underestimation of the tip indentation modulus and in the overestimation of the tip-sample contact radius. We propose a new approach to the calibration and data analysis, where lateral forces and cantilever inclination are neglected (which simplifies the calculations) and the tip parameters are assumed as fictitious.
Keywords
Contact resonance atomic force microscopy (CR-AFM) , Tip calibration , Atomic force acoustic microscopy (AFAM) , lateral force , Elastic modulus
Journal title
Astroparticle Physics
Record number
2043933