Title of article
The use of a central beam stop for contrast enhancement in TEM imaging
Author/Authors
Zhang، نويسنده , , Chao and Xu، نويسنده , , Qiang and Peters، نويسنده , , Peter J. and Zandbergen، نويسنده , , Henny، نويسنده ,
Pages
7
From page
200
To page
206
Abstract
Dark field TEM imaging using a stop of the central beam (DF-000) is reported. It is shown that a strong enhancement in the contrast can be obtained for graphene as example of weak phase object and endocytic multivescilar body as example of an unstained biological sample. No charging or significant contamination of the central beam stop is observed. For graphene, a resolution beyond 1 Å−1 was easily obtained. DF-000 imaging can be considered as a good and easy to use alternative of a phase plate.
Keywords
Center stop , Dark field imaging , Biological TEM , Phase plate , Contrast enhancement
Journal title
Astroparticle Physics
Record number
2044247
Link To Document