• Title of article

    Correction of three-fold astigmatism for ultra-high-resolution TEM

  • Author/Authors

    Overwijk، نويسنده , , M.H.F. and Bleeker، نويسنده , , A.J. and Thust، نويسنده , , A.، نويسنده ,

  • Pages
    8
  • From page
    163
  • To page
    170
  • Abstract
    Due to the influence of three-fold astigmatism, the spatial resolution obtained on many modern high-resolution transmission electron microscopes is often not better than 0.2 nm. Although beam tilt can mask the effect of three-fold astigmatism when imaging periodic specimens, the correction of three-fold astigmatism is required for obtaining isotropic ultra-high resolution (≤0.15 nm). Three-fold astigmatism also hampers the correct alignment of the microscope, resulting in additional beam tilt and two-fold astigmatism, and consequently in a further loss of resolution. The amount of three-fold astigmatism of an instrument is found to be stable for periods well in excess of a year. After hardware correction through excitation of a magnetic hexapole in the objective lens stigmator unit of our Philips CM300ST FEG, we reduced the magnitude of the three-fold astigmatism from a value of 1.1 μm, for this particular instrument, to less than 0.1 μm, a value at which it allows at least 0.1-nm spatial resolution. Additionally, we show that the micrscope can now be properly aligned for ultra-high-resolution experiments using the standard manual alignment procedures.
  • Keywords
    ALIGNMENT , Instrumental control , High resolution transmission electron microscopy (HRTEM)