Title of article :
Design of an objective lens pole piece for a transmission electron microscope with a resolution less than 0.1 nm at 200 kV
Author/Authors :
Tsuno، نويسنده , , K and Jefferson، نويسنده , , D.A، نويسنده ,
Pages :
9
From page :
31
To page :
39
Abstract :
The limit of the chromatic aberration coefficient Cc of a 200 kV transmission electron microscope has been calculated numerically. The limit is 0.34 mm, if no limitation exists for the specimen insertion and machining of the pole piece. The corresponding image side maximum of the axial flux density is 7.8 T. The computed Cc is nearly equal to the theoretical minimum of 0. 33 mm obtained from Schiskeʹs equation (P Schiske Optik 13 (1956) 502). Under practical conditions, Cc=0.63 mm and the corresponding spherical aberration coefficient Cs=0.55 mm are obtained. The dimensions of the pole piece are the gap length 2S=1.2 mm, the bore radius b=0.2 and the top face radius D=0.3 mm. Axial magnetic field distributions of such a pole piece have two peaks in the gap, because of the small bore and the top face area of the pole piece compared with the gap length. The resolution of the information limit is estimated to be 0.07 nm, if the stability of the high voltage is below 1×10−7 and a field emission gun is used.
Journal title :
Astroparticle Physics
Record number :
2046722
Link To Document :
بازگشت