Title of article :
Room design for high-performance electron microscopy
Author/Authors :
Muller، نويسنده , , David A. and Kirkland، نويسنده , , Earl J. and Thomas، نويسنده , , Malcolm G. and Grazul، نويسنده , , John L. and Fitting، نويسنده , , Lena and Weyland، نويسنده , , Matthew، نويسنده ,
Pages :
8
From page :
1033
To page :
1040
Abstract :
Aberration correctors correct aberrations, not instabilities. Rather, as spatial resolution improves, a microscopeʹs sensitivity to room environment becomes more noticeable, not less. Room design is now an essential part of the microscope installation process. Previously ignorable annoyances like computer fans, desk lamps and that chiller in the service corridor now may become the limiting factors in the microscopes performance. We discuss methods to quantitatively characterize the instrumentʹs response to magnetic, mechanical, acoustical and thermal disturbances and thus predict the limits that the environment places on imaging and spectroscopy.
Keywords :
High-performance electron microscopy
Journal title :
Astroparticle Physics
Record number :
2048395
Link To Document :
بازگشت