Title of article :
Characterization of a direct detection device imaging camera for transmission electron microscopy
Author/Authors :
Milazzo، نويسنده , , Anna-Clare and Moldovan، نويسنده , , Grigore and Lanman، نويسنده , , Jason and Jin، نويسنده , , Liang and Bouwer، نويسنده , , James C. and Klienfelder، نويسنده , , Stuart and Peltier، نويسنده , , Steven T. and Ellisman، نويسنده , , Mark H. and Kirkland، نويسنده , , Angus I. and Xuong، نويسنده , , Nguyen-Huu، نويسنده ,
Pages :
4
From page :
741
To page :
744
Abstract :
The complete characterization of a novel direct detection device (DDD) camera for transmission electron microscopy is reported, for the first time at primary electron energies of 120 and 200 keV. Unlike a standard charge coupled device (CCD) camera, this device does not require a scintillator. The DDD transfers signal up to 65 lines/mm providing the basis for a high-performance platform for a new generation of wide field-of-view high-resolution cameras. An image of a thin section of virus particles is presented to illustrate the substantially improved performance of this sensor over current indirectly coupled CCD cameras.
Keywords :
Active pixel sensor , Direct detection device , MTF , DQE , CMOS detectors , Transmission electron microscopy
Journal title :
Astroparticle Physics
Record number :
2050150
Link To Document :