Title of article :
Quantification of metallic nanoparticle morphology on TiO2 using HAADF-STEM tomography
Author/Authors :
Sueda، نويسنده , , S. and Yoshida، نويسنده , , K. and Tanaka، نويسنده , , N.، نويسنده ,
Pages :
8
From page :
1120
To page :
1127
Abstract :
Electron tomography is applied to photocatalytic gold/titanium oxide and gold/silver/titanium oxide samples. In order to obtain a tilt series for the electron tomography measurement, high-angle annular dark-field scanning transmission electron microscopy (HAADF-STEM) is used under cryogenic conditions. Dedicated programs have been developed for measuring volume, surface area, thickness distribution and nearest-neighbour distance of metallic nanoparticles on samples. Using these quantification programs, the 3D morphology of gold and silver nanoparticles is accurately characterized. We paid particular attention to the quantitative measurement of surface area. The measurement error of the method and appropriate magnification are defined using spherical nanoparticle models. We measured the 3D morphology of gold nanoparticles supported on titanium oxide (total volume=6.5×105 [nm3], surface area=1.4×105 [nm2], and average nearest-neighbour distance=40 [nm]).
Keywords :
Electron tomography , Metallic nanoparticles , Gold , Titanium oxide , Scanning transmission electron microscopy , STEM
Journal title :
Astroparticle Physics
Record number :
2050303
Link To Document :