Title of article :
Practical methods for the measurement of spatial coherence—A comparative study
Author/Authors :
Maunders، نويسنده , , C. P. Dwyer، نويسنده , , C. and Tiemeijer، نويسنده , , P.C. and Etheridge، نويسنده , , J.، نويسنده ,
Pages :
10
From page :
1437
To page :
1446
Abstract :
Two new methods for the measurement of transverse spatial coherence in a transmission electron microscope (TEM) are developed and applied to measure the spatial coherence in a field emission gun TEM. Measurements are made under different illumination and operating conditions, illustrating the effect of these conditions on the spatial coherence. The relative merits and limitations of these methods are discussed and compared, together with the previously described “Ronchigram” method.
Keywords :
COHERENCE , Source size , STEM , HRTEM
Journal title :
Astroparticle Physics
Record number :
2050905
Link To Document :
بازگشت