• Title of article

    Calibration of AFM cantilever spring constants

  • Author/Authors

    Gibson، نويسنده , , Christopher T. and Weeks، نويسنده , , Brandon L. and Abell، نويسنده , , Chris and Rayment، نويسنده , , Trevor and Myhra، نويسنده , , Sverre، نويسنده ,

  • Pages
    6
  • From page
    113
  • To page
    118
  • Abstract
    In this paper we present two simple, reliable and readily applicable methods for calibrating cantilevers and measuring the thickness of thin gold films. The spring constant calibration requires knowledge of the Youngʹs modulus, density of the cantilever and resonant frequency. The thickness of thin gold layers was determined by measuring changes in the resonant frequency and Q-factor of beam shaped AFM cantilevers before and after coating. chniques for measuring the spring constant and thin film thickness provide accuracy on the order of 10–15%.
  • Keywords
    Cantilever , Q-factor , Resonant frequency and spring constant
  • Journal title
    Astroparticle Physics
  • Record number

    2051497