Title of article
A new approach to scatter correction in SPECT images
Author/Authors
سفري، محسن حاجي زاده نويسنده Medical Physics Research Center, School of Medicine, Mashhad University of Medical Sciences, Saffar, Mohsen Hajizadeh , علومي، شبنم نويسنده دانشكده پزشكي-دانشگاه علوم پزشكي مشهد , , كنول، پيتر نويسنده Department of Nuclear Medicine, Wilhelminenspital, Vienna, Austria Knoll, Peter , تالشي، هادي نويسنده Department of Medical Physics Taleshi, Hadi
Issue Information
فصلنامه با شماره پیاپی 39 سال 2013
Pages
7
From page
19
To page
25
Abstract
Introduction: Scattered photon is one of the main defects that degrade the quality and quantitative accuracy of nuclear
medicine images. Accurate estimation of scatter in projection data of SPECT is computationally extremely demanding for
activity distribution in uniform and non-uniform dense media.
Methods: The objective of this paper is to develop and validate a scatter correction technique that use an accurate analytical
model based on Klein_Nishina scatter equation and compare Klein_Nishina scatter estimation with triple energy window. In
order to verify the proposed scattering model several cylindrical phantoms were simulated. The linear source in the
cylindrical Phantoms was a hot rod filled with 99mTc. K factor defines as the ratio of scatter resulting from MC simulation
to scatter estimated from Klein_Nishina formula. Also a SPECT/CT scan of the image quality phantom was acquired. Row
data were transferred to a PC computer for scatter estimation & processing of the images using MLEM iterative algorithm in
MATLAB software.
Results: The scatter and attenuation compensated images by the proposed model had better contrast than uncorrected and
only attenuation corrected images. The K-factors that used in proposed model doesn’t vary with different activities &
diameters of linear source and they’re just a function of depth and composition of pixels.
Conclusion: Based on Mont Carlo simulation data, the K_N formula that used in this study demonstrates better estimation
of scattered photons than TEW. Proposed scattered correction algorithm will improve 52.3% in the contrast of the attenuated
corrected images of image quality phantom.