Title of article :
Development of potentiometric creatinine-sensitive biosensor based on ISFET and creatinine deiminase immobilised in PVA/SbQ photopolymeric membrane
Author/Authors :
Soldatkin، نويسنده , , A.P and Montoriol، نويسنده , , J and Sant، نويسنده , , W and Martelet، نويسنده , , C and Jaffrezic-Renault، نويسنده , , N، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
5
From page :
75
To page :
79
Abstract :
A creatinine-sensitive biosensor was developed using ion-sensitive field-effect transistors (ISFETs) as transducers and immobilised creatinine deiminase (CD) as bioselective element. CD was immobilised by UV photopolymerisation in poly(vinyl alcohol) containing styrylpyridinium (PVA/SbQ) membrane on the dielectric gate of the ISFET transducer. veloped ENFETs demonstrated a dependence of the sensor sensitivity on NaCl and buffer concentration. Minimal detection limit for creatinine determination in a model solution containing 144 mM NaCl and 5% bovine serum albumin (BSA), pH 7.4, was about 20 μM. Biosensor responses were quite reproducible and stable during continuous work at least for 25 h, and the relative standard deviation of the sensor response was approximately 3% (n=45, for creatinine concentration of 0.1 mM). It was shown that creatinine-sensitive ENFETs demonstrated excellent storage stability for more than 6 months when kept dry at 4–6 °C.
Keywords :
PVA/SbQ photopolymer , ISFET , enzymes , Creatinine deiminase , Creatinine analysis , Biosensor , ENFET
Journal title :
Materials Science and Engineering C
Serial Year :
2002
Journal title :
Materials Science and Engineering C
Record number :
2097855
Link To Document :
بازگشت