• Title of article

    Characterization of a trap pyroelectric sensor noise and improvement of the measurement range

  • Author/Authors

    Mellouki، نويسنده , , I. and Sifi، نويسنده , , N. and Touayar، نويسنده , , O. and Abroug، نويسنده , , S. El Yacoubi a، نويسنده , , N.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    3
  • From page
    314
  • To page
    316
  • Abstract
    In this paper, a characterization of trap pyroelectric sensor noise was investigated, by adopting and characterizing a measuring equipment of particular design. The characterization of the background noise of this sensor shows a value of the Noise Equivalent Power (N.E.P) of about 0.146 μW/Hz1/2 which is compared with the results available in the literature. A characterization of the sensor N.E.P using a laser diode in the range of powers going from 100 μW to 40 mW is also presented. This study revealed that the considered trap sensor can provide an absolute measurement of the powers laser starting from 100 μW.
  • Keywords
    Noise characterization , Pyroelectric sensor , Noise equivalent power , Laser absolute power measurement
  • Journal title
    Materials Science and Engineering C
  • Serial Year
    2006
  • Journal title
    Materials Science and Engineering C
  • Record number