    Stiffness analysis of tibia-implant system under cyclic loading

  • Author/Authors

    Tschegg، نويسنده , , Elmar K. and Herndler، نويسنده , , Stefan and Weninger، نويسنده , , Patrick and Jamek، نويسنده , , Michael and Stanzl-Tschegg، نويسنده , , Stefanie and Redl، نويسنده , , Heinz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    1203
  • To page
    1208
  • Abstract
    A testing method has been developed to characterize stiffness and permanent deformation of bone-implant systems. The system consists of an artificial tibia with simulated distal fracture stabilized by an unreamed intramedullary nail. This system was loaded with three different sequences at 2 Hz, each consisting of 40,000 sinusoidal cycles, simulating clinical relevant loading conditions. Evaluation of the results showed a stiffness of 2782 N/mm with a standard deviation of 311 N/mm and a permanent deformation of 0.64 mm with a standard deviation of 0.21 mm. The locking screws broke exclusively during the third loading sequence starting with the most proximal of the distal screws. The study provides a standard technique for biomechanical testing and a comparison of different bone/implant-systems avoiding the variability of cadaver bone tests.
  • Keywords
    Polymer-composite bone , Distal tibia fracture , Bone-implant-system , Fatigue loading , Stiffness , permanent deformation