Title of article :
Iodide selective membrane electrodes based on a Molybdenum–Salen as a neutral carrier
Author/Authors :
Ghanei-Motlagh، نويسنده , , Masoud and Taher، نويسنده , , Mohammad Ali and Ahmadi، نويسنده , , Kyoumars and Sheikhshoaie، نويسنده , , Iran، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
7
From page :
1625
To page :
1631
Abstract :
A new polymeric membrane electrode (PME) and a coated platinum disk electrode (CPtE) based on Schiff base complex of Mo(VI) as a suitable carrier for I− ion were described. The influence of membrane composition, pH and possible interfering anions were investigated on the response properties of the electrodes. The electrodes exhibited a Nernstian slope of 63.0 ± 0.5 (CPtE) and 60.3 ± 0.4 (PME) mV decade− 1 in I− ion over a wide concentration range from 7.9 × 10− 7 to 1.0 × 10− 1 M for CPtE and 9.1 × 10− 6 to 1.0 × 10− 1 M I− for PME. The potentiometric response of the electrodes was independent of the pH of the test solution in the pH range 2.0–8.5 with a fast response time (< 10 s). The process of transfer of iodide across the membrane interface was investigated by use of the AC impedance technique. The proposed sensors were successfully applied to direct determination of iodide in samples containing interfering anions, waste water and as indicator electrodes in precipitation titrations.
Keywords :
Coated platinum disk electrode , Potentiometry , Schiff base complex , PVC Membrane Electrode , Iodide selective electrode