Title of article :
Development of a flow injection analysis (FIA) immunosensor for the detection of Escherichia coli
Author/Authors :
Bouvrette، نويسنده , , P. and Luong، نويسنده , , J.H.T.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
9
From page :
129
To page :
137
Abstract :
A flow injection immunoanalysis (FIA) system has been developed for the detection of Escherichia coli in artificially contaminated food samples. Anti-E. coli antibodies were covalently immobilized onto porous aminopropyl glass beads via glutaraldehyde activation to form an immunoreactor. After adsorption of the cells onto anti-E. coli antibody bound glass beads, 4-methylumbelliferyl-β-d-glucuronide was injected into the system which was then hydrolyzed by the adsorbed E. coli cells containing β-d-glucuronidase, an enzyme which is very specific to E. coli and to a few other strains of Shigella. Fluorescent 4-methylumbelliferone released from the enzymatic reaction was then detected by a fluorometer. Owing to the specificity of the antibody towards E. coli, the FIA system was very selective for detection of E. coli whereas Shigella boydii, another GUD-positive bacterium, did not give any response. The FIA system was successfully used for detecting as low as 5 × 107 CFU/ml E. coli in less than 30 min and was reusable for at least 300 repeated assays. The immunoreactor yielded reproducible results during 3 months of experimentation if stored overnight at 4 °C in carrier buffer containing 0.05 to 0.25% Tween 20.
Keywords :
Escherichia coli , Immunosensor , Anti-E. coli antibody , fluorescence , 4-Methylumbelliferone , Flow injection analysis
Journal title :
International Journal of Food Microbiology
Serial Year :
1995
Journal title :
International Journal of Food Microbiology
Record number :
2107034
Link To Document :
بازگشت