Title of article
Problems in temperature control performing in situ investigations with the scanning force microscope
Author/Authors
Kipp، نويسنده , , S. and Lacmann، نويسنده , , R. and Schneeweiss، نويسنده , , M.A.، نويسنده ,
Issue Information
دوماهنامه با شماره پیاپی سال 1995
Pages
3
From page
333
To page
335
Abstract
Up to now, various publications dealing with in situ scanning force microscopy without any temperature control have been published. In this paper it is demonstrated that the temperature increase in the sample chamber due to the use of a diode laser in the detection unit cannot be neglected. The effect was quantified as 3–4 K per hour in the non-thermostatted SFM cell and by the recirculation of thermostatted liquid limited to 0.6 k.