• Title of article

    Quantitative amplitude and phase contrast imaging in a scanning transmission X-ray microscope

  • Author/Authors

    Hornberger، نويسنده , , Benjamin and Feser، نويسنده , , Michael and Jacobsen، نويسنده , , Chris، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2007
  • Pages
    12
  • From page
    644
  • To page
    655
  • Abstract
    Phase contrast in X-ray imaging provides lower radiation dose, and dramatically higher contrast at multi-keV photon energies when compared with absorption contrast. We describe here the use of a segmented detector in a scanning transmission X-ray microscope to collect partially coherent bright field images. We have adapted a Fourier filter reconstruction technique developed by McCallum, Landauer and Rodenburg to retrieve separate, quantitative maps of specimen phase shift and absorption. This is demonstrated in the imaging of a germanium test pattern using 525 eV soft X-rays.
  • Keywords
    X-ray microscopy , Phase contrast , Phase recovery , Configured detector , Segmented detector
  • Journal title
    Ultramicroscopy
  • Serial Year
    2007
  • Journal title
    Ultramicroscopy
  • Record number

    2156931