• Title of article

    Developing a Confocal Acoustic Holography Microscope for non-invasive 3D temperature and composition measurements

  • Author/Authors

    Herring، نويسنده , , Rodney A. and Jacquemin، نويسنده , , Peter and Sawicka، نويسنده , , Barbara D. and Atalick، نويسنده , , Stefan، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2009
  • Pages
    7
  • From page
    830
  • To page
    836
  • Abstract
    A Confocal Acoustic Holography Microscope (CAHM) has been designed, simulated and partially verified experimentally to take holograms for non-invasive, three-dimensional measurements of a specimenʹs refractive indices from one view point. The designed and simulated prototype CAHM used a frequency of 2.25 MHz and measured sound speed changes of 16 m/s, temperature changes of 5 °C and had a spatial resolution of 660 μm. With future improvements utilizing the latest technologies such as two-dimensional array detectors, Micro-Electro-Mechanical Systems (MEMS), and acoustic lenses, resolutions of 1 m/s, 0.5 °C, and 150 μm are expected. The CAHM is expected to have many useful applications, including non-invasive mass and heat transfer measurements in fluids and materials and as a medical diagnostic tool to non-intrusively visualize compositions and temperatures within the human body.
  • Keywords
    confocal , Holography , Acoustic beams , Microscopy
  • Journal title
    Ultramicroscopy
  • Serial Year
    2009
  • Journal title
    Ultramicroscopy
  • Record number