Title of article :
Interlaboratory round robin on cantilever calibration for AFM force spectroscopy
Author/Authors :
te Riet، نويسنده , , Joost and Katan، نويسنده , , Allard J. and Rankl، نويسنده , , Christian and Stahl، نويسنده , , Stefan W. and van Buul، نويسنده , , Arend M. and Phang، نويسنده , , In Yee and Gomez-Casado، نويسنده , , Alberto and Schِn، نويسنده , , Peter R.W. Gerritsen، نويسنده , , Jan W. and Cambi، نويسنده , , Alessandra and Rowan، نويسنده , , Alan E. and Vancso، نويسنده , , G. Julius and Jonk، نويسنده ,
Issue Information :
دوماهنامه با شماره پیاپی سال 2011
Pages :
11
From page :
1659
To page :
1669
Abstract :
Single-molecule force spectroscopy studies performed by Atomic Force Microscopes (AFMs) strongly rely on accurately determined cantilever spring constants. Hence, to calibrate cantilevers, a reliable calibration protocol is essential. Although the thermal noise method and the direct Sader method are frequently used for cantilever calibration, there is no consensus on the optimal calibration of soft and V-shaped cantilevers, especially those used in force spectroscopy. Therefore, in this study we aimed at establishing a commonly accepted approach to accurately calibrate compliant and V-shaped cantilevers. In a round robin experiment involving eight different laboratories we compared the thermal noise and the Sader method on ten commercial and custom-built AFMs. We found that spring constants of both rectangular and V-shaped cantilevers can accurately be determined with both methods, although the Sader method proved to be superior. Furthermore, we observed that simultaneous application of both methods on an AFM proved an accurate consistency check of the instrument and thus provides optimal and highly reproducible calibration. To illustrate the importance of optimal calibration, we show that for biological force spectroscopy studies, an erroneously calibrated cantilever can significantly affect the derived (bio)physical parameters. Taken together, our findings demonstrated that with the pre-established protocol described reliable spring constants can be obtained for different types of cantilevers.
Keywords :
Cantilever calibration , atomic force microscopy , Round robin experiment , Spring Constant , AFM force spectroscopy
Journal title :
Ultramicroscopy
Serial Year :
2011
Journal title :
Ultramicroscopy
Record number :
2158447
Link To Document :
بازگشت