Abstract :
Precession electron diffraction has seen a fast increase in its adoption as a technique for solving crystallographic structures as well as an alternative to conventional selected-area and converged-beam diffraction methods. One of the key issues of precession is the pivot point alignment, as a stationary apparent beam does not guarantee a fixed pivot point. A large precession tilt angle, along with pre-field and post-field misalignment, induces shift in the image plane. We point out here that the beam should be aligned to the pre-field optic axis to keep the electron illumination stationary during the rocking process. A practical alignment procedure is suggested with the focus placed on minimizing the beam wandering on the specimen, and is demonstrated for a (110)-oriented silicon single crystal and for a carbide phase (∼20 nm in size) within a cast cobalt–chromium–molybdenum alloy.
Keywords :
Precession electron diffraction , CoCrMo alloy , ALIGNMENT , M23C6 carbide