Title of article :
Polymeric membrane neodymium(III)-selective electrode based on 11,13-diaza-4,7,12-trioxo-2(3),8(9)-dibenzoyl- cyclotetridecane-1,11-diene
Author/Authors :
Chandra، نويسنده , , Sulekh and Singh، نويسنده , , Dev Raj، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
4
From page :
107
To page :
110
Abstract :
We found that 11, 13-diaza-4, 7, 12-trioxo-2(3), 8(9)-dibenzoyl-cyclotetridecane-1, 11-diene (DATODBCT) can be used as an excellent ionophore in the construction of a novel neodymium(III) poly(vinylchloride) (PVC)-based membrane sensor. A membrane composition of 29% poly(vinylchloride), 59% dibutylphthalate (DBP), 9% DATODBCT and 3% sodiumtetrakis (p-chlorophenyl) borate (NaTpClPB), led to the optimum results. The Nd(III)-selectivity of the sensor, is relatively better as compared to a large number of lanthanide metal ions, such as lanthanum, gadolinium, samarium, dysprosium, praseodymium and ytterbium ions. The sensor response is Nernstian (with slope of 19.4 ± 0.3 mV per decade for the triply charged ion) over a wide concentration range (1.0 × 10−8 to 1.0 × 10−1 mol L−1) with a detection limit of 8.0 × 10−7 mol L−1, a relatively fast response time, in the whole concentration range (<15 s), and a considerable life time at least for seven weeks in the pH range of 3.0–7.0.
Keywords :
Nd3+-ion selective electrode , PVC Membrane , Potentiometry , Nernstian slope