Title of article :
Energy dependence of gamma detection efficiency for SED R-12
Author/Authors :
Sarkar، نويسنده , , R. and Chatterjee، نويسنده , , B.K. and Roy، نويسنده , , S.C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
4
From page :
203
To page :
206
Abstract :
Superheated emulsion detector consists of droplets of two metastable states, one with a much shorter life-time (second metastable state) than the other (normal metastable state). The SED made from R-12 (CCl2F2) is sensitive to gamma rays near 40 °C. Gamma detection sensitivity of superheated R-12 increases monotonically beyond this threshold temperature. The radiation induced nucleation frequency of normal superheated drops of R-12 is studied in the temperature range 45–58 °C for gamma rays of energy 59.54 and 662 keV obtained from 241Am and 137Cs sources, respectively. It is observed that the radiation induced nucleation frequency and hence the gamma detection efficiency increase with temperature and with the energy of the gamma rays.
Keywords :
Two states , Radiation induced nucleation , Superheated emulsion detector , Detection efficiency
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2010
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2170156
Link To Document :
بازگشت