Title of article
Application of digital zero-crossing technique for neutron–gamma discrimination in liquid organic scintillation detectors
Author/Authors
Nakhostin، نويسنده , , M. and Walker، نويسنده , , P.M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
4
From page
498
To page
501
Abstract
An algorithm for digital implementation of the zero-crossing method for n/γ discrimination in liquid organic scintillators is described. The method exhibits good performance at low energies and requires little computational effort, which makes it suitable for compact real-time neutron detectors.
Keywords
Liquid scintillator , Digital pulse-shape discrimination , neutron detection
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2010
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2170419
Link To Document