• Title of article

    Temperature characterization of scintillation detectors using solid-state photomultipliers for radiation monitoring applications

  • Author/Authors

    Tur، نويسنده , , Clarisse and Solovyev، نويسنده , , Vladimir and Flamanc، نويسنده , , Jérémy، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    8
  • From page
    351
  • To page
    358
  • Abstract
    We have characterized two state-of-the-art solid-state photomultipliers, one by SensL, the other by Hamamatsu, coupled to scintillators by Saint-Gobain Crystals in the −25 to +50 °C temperature range. At room temperature, the energy resolution at 661.6 keV measured with both detectors is worse than the resolution obtained when the crystals are coupled to a regular photomultiplier tube. Both the pulse height and pulse height resolution of the 661.6 keV gamma rays in the 137Cs spectrum vary strongly with temperature. The noise threshold determined from the 22Na spectrum increases quadratically as the temperature is increased to well above 100 keV at +50 °C for both detectors.
  • Keywords
    Multi-pixel photon counter , MPPC , Solid-state photomultiplier , Geiger-mode avalanche photodiode array , SSPM , G-APD array , scintillation detector , Temperature characterization , Silicon PhotoMultiplier , SPM , SiPM
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2010
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2170937