Title of article :
Geometry effects in well-type Ge-detectors due to the introduction of high-Z linings
Author/Authors :
Gelsema، نويسنده , , S.J. and Blaauw، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
6
From page :
410
To page :
415
Abstract :
In well-type detectors and other highly efficient geometries, the combined effects of sample self-absorption involving low-energy gamma-rays and coincidence summing of a low- and a high-energy photon often influence the entire gamma-spectrum of a radionuclide. For well-type detectors, these effects are well understood and it should be possible to correct for these effects analytically. Another possibility to deal with these effects is to prevent the low-energy gamma-rays from reaching the detector by lining the well with a high-Z material, as was suggested in the past. sent a series of measurements which indicates that the problems resulting from sample self-absorption and coincidence summing should be treated analytically rather than by attempts to eliminate the effects by means of a lining.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2171233
Link To Document :
بازگشت