Title of article :
A new low background gamma spectrometer equipped with an anti-cosmic device
Author/Authors :
J. Laurec، نويسنده , , J. and Blanchard، نويسنده , , X. and Pointurier، نويسنده , , F. and Adam، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
6
From page :
566
To page :
571
Abstract :
A well type Ge detector and a 48 percent relative efficiency coaxial Ge detector are used with cosmic ray anticoincidence shielding (plastic scintillating plates). Although these devices have lower performance than those placed in an underground laboratory, their advantage is that they can be used in a normal laboratory environment. The background is reduced by 4 or 5 in the whole 30 keV to 2.7 MeV energy range and by 5 to 10 in the energy range below 500 keV. Under these conditions, the count rates are smaller than 10−4 count s−1 keV−1.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2171537
Link To Document :
بازگشت