Title of article :
Calibration of the energy analyzing magnet of the NSC Pelletron using a differential time of flight technique
Author/Authors :
Sarkar، نويسنده , , A. and Ghosh، نويسنده , , S. and Joshi، نويسنده , , R. and Kanjilal، نويسنده , , D. and Roy، نويسنده , , Marcos A. and Bhowmik، نويسنده , , R.K.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
351
To page :
354
Abstract :
A differential time of flight technique was tried using two cavity resonators (as timing devices) for the calibration of the energy analyzing magnet of the pelletron accelerator at NSC. This method is unique as it compensates for the intrinsic delays for the detectors used in other TOF experiments. The experimentally measured value of the radius of curvature of the analyzing magnet is 1808 mm which is comparable to the value obtained from the energy calibration by compound nuclear resonance in the 12C (p,p′) reaction.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2172171