Title of article :
A study of the timing characteristics of scintillator counters with fine-mesh photomultiplier tubes
Author/Authors :
Foucher، نويسنده , , M and Herndon، نويسنده , , M and Jawahery، نويسنده , , H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
6
From page :
57
To page :
62
Abstract :
We report on a study of the precision of Time-Of-Flight measurements using plastic scintillator counters with Fine Mesh Photomultiplier Tubes (FM PMT). This work was part of an investigation of several techniques considered for hadron identification in the B factory detector (BaBar) at SLAC. In cosmic ray tests of prototype counters we achieved a time resolution of 91 ps with a 3.0 m long counter and 70 ps with a 1.0 m long counter.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2172601
Link To Document :
بازگشت