Calibration of plastic phoswich detectors for charged particle detection
Author/Authors :
Fox، نويسنده , , D and Bowman، نويسنده , , D.R and Ball، نويسنده , , G.C and Galindo-Uribarri، نويسنده , , A and Hagberg، نويسنده , , E and Horn، نويسنده , , D and Beaulieu، نويسنده , , L and Larochelle، نويسنده , , Y، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
7
From page :
63
To page :
69
Abstract :
The response of an array of plastic phoswich detectors to ions of 1 ≤ Z ≤ 18 has been measured from EA = 12 to 72 MeV. The detector response has been parameterized by a three parameter fit which includes both quenching and high energy delta-ray effects. The fits have a mean variation of ≤4% with respect to the data.