• Title of article

    Design of a divergence and alignment indicator

  • Author/Authors

    Brenizer Jr.، نويسنده , , J.S. and Raine، نويسنده , , Da-Kuan Gao، نويسنده , , J. and Chen، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    5
  • From page
    58
  • To page
    62
  • Abstract
    The divergence and alignment indicator (DAI) is an extension of the ASTM E803 L/D thermal neutron radiography L/D device that allows the user to determine both the beam centerline and the beam divergence. The DAI was made using aluminium plate and rods, and incorporated cadmium wire for contrast. Circular symmetry was utilized to simplify manufacture. The DAI was placed with the five posts against the film cassette or radioscopic imaging device in the physical center of the beam. The DAI was perpendicular to the selected beam radius when the front and back center Cd wire images overlap. The degree of misalignment was indicated by their image positions. After the DAI was aligned, analysis of the cadmium wire “+” image spacing yielded the beam divergence. The DAI was tested in a neutron beam which has an L/D of 30 but a small degree of divergence. The DAI was also imaged using an X-ray source. The point source predictions of Cd wire image locations showed good agreement with those measured from the X-ray radiograph. The neutron radiographic locations could be predicted using the point source equations, even though the neutron beam was a complex distributed source.
  • Keywords
    Divergence angle , ALIGNMENT , Beam characterization , Neutron radiography
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1996
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2173121