Title of article :
Portable high energy gamma ray imagers
Author/Authors :
Guru، نويسنده , , S.V and He، نويسنده , , Z and Wehe، نويسنده , , D.K and Knoll، نويسنده , , G.F and Redus، نويسنده , , R.H and Squillante، نويسنده , , M.R، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
8
From page :
612
To page :
619
Abstract :
To satisfy the needs of high energy gamma ray imagers for industrial nuclear imaging applications, three high energy gamma cameras are presented. The RMD-Pinhole camera uses a lead pinhole collimator and a segmented BGO detector viewed by a 3 in. square position sensitive photomultiplier tube (PSPMT). This pinhole gamma camera displayed an energy resolution of 25.0% FWHM at the center of the camera at 662 keV and an angular resolution of 6.2° FWHM at 412 keV. The fixed multiple hole collimated camera (FMCC), used a multiple hole collimator and a continuous slab of NaI(Tl) detector viewed by the same PSPMT. The FMCC displayed an energy resolution of 12.4% FWHM at 662 keV at the center of the camera and an angular resolution of 6.0° FWHM at 412 keV. The rotating multiple hole collimated camera (RMCC) used a 180° antisymmetric rotation modulation collimator and CsI(Tl) detectors coupled to PIN silicon photodiodes. The RMCC displayed an energy resolution of 7.1% FWHM at 662 keV and an angular resolution of 4.0° FWHM at 810 keV. The performance of these imagers is discussed in this paper.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2173394
Link To Document :
بازگشت