Title of article
Timing performances and edge effects of detectors worked from 6-in. silicon slices
Author/Authors
Aiello، نويسنده , , S. and Anzalone، نويسنده , , A. and Cardella، نويسنده , , G. and Cavallaro، نويسنده , , Sl. and De Filippo، نويسنده , , E. and Di Pietro، نويسنده , , A. and Feminَ، نويسنده , , S. and Geraci، نويسنده , , M. and Giustolisi، نويسنده , , F. and Guazzoni، نويسنده , , P. and Iacono-Manno، نويسنده , , M. and Lanzalone، نويسنده , , G. and Lanzanَ، نويسنده , , G. and Lo Nigro، نويسنده , , S. and Musumarra، نويسنده , , A، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
5
From page
306
To page
310
Abstract
Prototypes of new passivated implanted planar silicon detectors, obtained for the first time from 6 in. silicon slices, have been tested. The time and energy resolutions have been studied as a function of the type and energy of the detected particles, in order to test the performances of these detectors for time of flight measurements in the Chimera project. Some problems arising from edge effects observed in double-pad detectors have been solved by using a guard ring.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1997
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2174758
Link To Document