Author/Authors :
Bialkowski، نويسنده , , J. and Bühler، نويسنده , , P. and Henneck، نويسنده , , R. and Martini، نويسنده , , R. and Mchedlishvili، نويسنده , , A. and Schlumpf، نويسنده , , N. and Schoeps، نويسنده , , W. and Zehnder، نويسنده , , A. and Siegmund، نويسنده , , O.، نويسنده ,
Abstract :
The Extreme UV Imaging Detector EUVID consists of a compact “micro-channel-plate-wedge-and-strip anode” imaging system and corresponding data read-out electronics with 32 MB of memory. EUVID is suited for imaging in the wavelength range 20 ≤ λ ≤ 1200 Å. The performance of this space-qualified, 2.5 kg detector system was measured in the FUV. We obtained a spatial resolution of about 70 μm over the full detector area of ∼ 7 cm2. Image distortions are small and can easily be corrected for. Detector efficiency depends on the choice of photocathode and ranges between 50% in the FUV up to 90% in the EUV.