Title of article
The dynamic range of X-ray imaging with image plates
Author/Authors
Thoms، نويسنده , , M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
4
From page
437
To page
440
Abstract
The dynamic range of X-ray detection with image plates is investigated. A dynamic range larger than 108 in X-ray dose is experimentally observed using the Fastscan detector, which allows a wide range of sensitivity settings. At X-ray doses lower than 0.1 Gy a proportionality between photo-stimulated luminescence (PSL) intensity and X-ray dose is found, while at higher doses the PSL intensity starts to saturate. This saturation effect can be attributed to the coloration of heavily X-rayed image-plates by radiation-induced F-centers.
Keywords
Imaging , photostimulated luminescence , Storage phosphor , x-ray detection , X-Ray , BaFBr , color center , detector , dynamic range , F-center , Image read out , Photo-stimulation , Image plate , EU
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1997
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2175403