Title of article :
Characterising a Si(Li) detector element for the SIXA X-ray spectrometer
Author/Authors :
Tikkanen، نويسنده , , T. and Kraft، نويسنده , , S. and Scholze، نويسنده , , F. and Thornagel، نويسنده , , R. and Ulm، نويسنده , , G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
The detection efficiency and response function of a Si(Li) detector element for the SIXA spectrometer have been determined in the 500 eV to 5 keV energy range using synchrotron radiation emitted at a bending magnet of the electron storage ring BESSY, which is a primary radiation standard. The agreement between the measured spectrum and the model calculation is better than 2%.
Keywords :
X-ray response , Spectral line shape , Si(Li) detectors , X-ray spectrometers , detector calibration
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A