Title of article
Position-sensitive gas detectors: MWPCs and their gifted descendants
Author/Authors
Fourme، نويسنده , , Roger، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
11
From page
1
To page
11
Abstract
On the basis of concepts used in conventional multi-wire proportional chambers and drift chambers, another generation of position-sensitive gaseous detectors has matured for a variety of applications. Several limitations of multi-wire chambers have improved, sometimes by several orders of magnitude. The basis features and performance of some current devices are discussed: standard symmetric MWPC, MSGC, asymmetric structures (including MWPC, MICROMEGAS, CAT and MGC), secondary electron emission gaseous detectors and gas proportional scintillation detectors.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1997
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number