Title of article :
Tests of a multianode PMT for the HERA-B RICH
Author/Authors :
Kri?an، نويسنده , , P and Korpar، نويسنده , , S and Pestotnik، نويسنده , , R and Stari?، نويسنده , , M and Stanovnik، نويسنده , , A and Michel، نويسنده , , E. and Oehser، نويسنده , , C and Schmidt-Parzefall، نويسنده , , W and Schwarz، نويسنده , , A and Hamacher، نويسنده , , T and Broemmelsiek، نويسنده , , D and Pyrlik، نويسنده , , J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
8
From page :
27
To page :
34
Abstract :
A new multianode photomultiplier tube, type H6568 manufactured by Hamamatsu company, has been tested in view of using it for measuring Cherenkov rings in the HERA-B experiment. The results of measurements of single-photoelectron pulse-height distributions, uniformity of response, cross-talk, photon detection efficiency, rate dependence of the gain and ageing will be presented and discussed. In addition, the response of the PMT to scintillations of perfluorobutane (C4F10), argon and air has been measured.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1997
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2176046
Link To Document :
بازگشت