Title of article :
Precision measurement of the magnetic field in terms of the free-proton NMR frequency
Author/Authors :
Fei، نويسنده , , Xiang and Hughes، نويسنده , , V.W. and Prigl، نويسنده , , Ralf، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
We have developed a NMR standard probe using a spherical pure water sample of 1 cm diameter to determine the absolute magnetic field B in terms of the free-proton NMR frequency fp with an accuracy of 3.4 × 10−8. Our standard probe can be used conveniently to calibrate other probes in the field range from 1.45 to 1.7 T and can readily be employed over a much wider field range. The probe design and the tests carried out to verify its precision and accuracy are presented.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A