Title of article :
Gas-silicon detector telescope for charged particle spectroscopy
Author/Authors :
Honkanen، نويسنده , , A and Oinonen، نويسنده , , M and Eskola، نويسنده , , K and Jokinen، نويسنده , , A and ؤystِ، نويسنده , , J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
9
From page :
217
To page :
225
Abstract :
A gas-silicon detector telescope for charged particle spectroscopy has been constructed and tested. The lower detection limits were determined to be 155 keV for protons, 180 keV for deuterons and 350 keV for alpha particles. Typical energy resolution of the telescope measured for beta-delayed protons is 20 keV. Time resolution for the signals of the telescope was measured to be less than 10 ns. Examples of using the detector telescope in detection of beta-delayed proton activities are presented.
Keywords :
Low-energy charged particle detector telescope , Proportional counter , Ion-implanted Si-detector
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1997
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2176228
Link To Document :
بازگشت