Title of article :
Successful test of a prototype hadron blind detector
Author/Authors :
Pisani، نويسنده , , R.P and Hemmick، نويسنده , , T.K and Chung، نويسنده , , H and Johnson، نويسنده , , S.C and Piazza، نويسنده , , T and Vongpaseuth، نويسنده , , T and Akopyan، نويسنده , , M، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Pages :
12
From page :
243
To page :
254
Abstract :
We have constructed and tested a threshold Cherenkov detector inspired by the conceptual design of Giomataris and Charpak (Nucl. Instr. and Meth. A 310 (1991) 589). The basic design consists of a gas radiator followed by a photosensitive wire chamber using a solid Csl photocathode. The photon detector lies directly in the particle path and is required to have single photo-electron sensitivity and yet be insensitive to the passage of a charged particle. We have altered the design by adding a thin LiF window. This window allows one to separate the radiator and avalanche volumes, giving greater flexibility in the gas choices. Our detector was operated with C2F6 as the radiator gas at lengths of 120, 80, and 40 cm. The avalanche gas was 95%He + 5% CH4. At 95% electron efficiency, the pion efficiency was 1333, 1150, and 1100 for each of the lengths listed above.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1997
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2176802
Link To Document :
بازگشت