Title of article :
Xe isotope detection and discrimination using beta spectroscopy with coincident gamma spectroscopy
Author/Authors :
Reeder، نويسنده , , P.L and Bowyer، نويسنده , , T.W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Beta spectroscopic techniques show promise of significant improvements for a beta–gamma coincidence counter that is part of a system for analyzing Xe automatically separated from air. The previously developed counting system for 131mXe, 133mXe, 133gXe, and 135gXe can be enhanced to give additional discrimination between these Xe isotopes by using the plastic scintillation sample cell as a beta spectrometer to resolve the conversion electron peaks. The automated system will be a key factor in monitoring the Comprehensive Test Ban Treaty.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A