Title of article :
Performance tests of the IAE dose equivalent meter in radiation field of high energy calibration facility at SPS-CERN
Author/Authors :
Rusinowski، نويسنده , , Z. and Golnik، نويسنده , , N.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
The performance of the IEA dose equivalent meter based on the REM-2 recombination chamber was tested in pulsed high energy radiation field at CERN-EC calibration facility. The device was working with its own monitoring circuit, and provided accurate and stable results, within 2% of statistical uncertainty.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A