Title of article :
Determining detector requirements for medical imaging applications
Author/Authors :
Clinthorne، نويسنده , , Neal H. and Ng، نويسنده , , Chor-yi and Strobel، نويسنده , , Joerg and Hua، نويسنده , , Chia-ho and LeBlanc، نويسنده , , James W. and Wilderman، نويسنده , , Scott J. and Rogers، نويسنده , , W.Leslie، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
7
From page :
501
To page :
507
Abstract :
The Cramér–Rao lower bound on estimator variance is used as a tool for assessing radiation detector requirements for nuclear medicine imaging. The classical bound for unbiased estimators is reviewed and its relationship to the well-known propagation-of-error formula is demonstrated. The uniform Cramér–Rao bound is described for situations where unbiased estimation is impractical such as in tomographic imaging. The bounds are used to evaluate the necessary detector requirements for a CsI(Tl)/Si photodiode scintillation camera, and to compare the performance of a Compton-scatter camera to that of a parallel-hole collimator/Anger camera system.
Keywords :
Noise-propagation , Cramér–Rao lower bound , Noise
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1998
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2178134