Title of article :
Electrical measurement of wire tension in a multi-wire drift chamber
Author/Authors :
Ohama، نويسنده , , John T. and Ishihara، نويسنده , , N. and Takeda، نويسنده , , S. and Okuma، نويسنده , , H. and Konno، نويسنده , , K.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
4
From page :
175
To page :
178
Abstract :
A new and simple technique has been developed to measure the wire tension of a multi-wire drift chamber, which is the prototype chamber of DCBA (Drift Chamber Beta-ray Analyzer) for double beta decay experiments. The present method uses two wires adjacent to each other in the chamber for applying AC and DC electric fields, and can easily detect a weak current signal at the resonance frequency corresponding to the wire tension. The wire tension has been measured with an accuracy of about 1%.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1998
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2178209
Link To Document :
بازگشت