• Title of article

    A simple circuit for background reduction in low level tritium activity measurement using liquid-scintillation counting

  • Author/Authors

    Vaidya، نويسنده , , P.P and Gurna، نويسنده , , R.K and Tabib، نويسنده , , J.D and Gopalakrishnan، نويسنده , , K.R، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1998
  • Pages
    3
  • From page
    188
  • To page
    190
  • Abstract
    The paper describes a new method for reduction of background counts in Liquid Scintillation Counting (LSC) using a simple circuit. With this circuit it was possible to reduce background counts in low level Tritium Counting to 8 cpm from an initial value of 18 cpm with a marginal degradation in detection efficiency. Use of the circuit can reduce the cost of passive shielding, which is generally used to reduce background counts to a low level.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1998
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2178718