Calibration and performance of the E706 lead and liquid-argon electromagnetic calorimeter
Author/Authors
Apanasevich، نويسنده , , L. and Bacigalupi، نويسنده , , J. and Baker، نويسنده , , W. and Ballocchi، نويسنده , , G. and Begel، نويسنده , , M. and Benson، نويسنده , , R. and Blusk، نويسنده , , Ronald S. and Bromberg، نويسنده , , C. and Brown، نويسنده , , D. and Chang، نويسنده , , P. and Choudhary، نويسنده , , B. and Chung، نويسنده , , W.H. and Barbaro، نويسنده , , L.de and DeSoi، نويسنده , , W. and D?ugosz، نويسنده , , W. and Dunlea، نويسنده , , J. and، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
19
From page
50
To page
68
Abstract
We report on the calibration and performance of a large lead liquid-argon electromagnetic calorimeter used by experiment E706 at Fermilab. The reconstructed π0 mass was used to calibrate the energy response of the calorimeter. The systematic uncertainty in the linearity and uniformity of the mean-energy response of the calorimeter after calibration was found to be less than ±0.5% for the sample of data analyzed. Detector characteristics, including sampling fluctuations and position resolution, are discussed.
Keywords
Sampling electromagnetic calorimeter , Calibration , direct photons