• Title of article

    Results from the BABAR electromagnetic calorimeter beam test

  • Author/Authors

    Barlow، نويسنده , , Roger J. and Bernet، نويسنده , , Roland and Bowdery، نويسنده , , Christopher K. and Jens Brose and Champion، نويسنده , , Theresa and Dahlinger، نويسنده , , Gerd and Dauncey، نويسنده , , Paul and Peter Eckstein and Eigen، نويسنده , , Gerald and Freytag، نويسنده , , Dietrich and Fry، نويسنده , , John and Geddes، نويسنده , , Neil I. and Hertzbach، نويسنده , , Stanley ، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    19
  • From page
    162
  • To page
    180
  • Abstract
    We present results from the PSI test beam run for the electromagnetic calorimeter of the BABAR experiment. A system of 25 CsI(Tl) crystals was tested with electrons and pions in the momentum range from 100 to 405 MeV/c. Results are presented on crystal light output, on coherent and incoherent noise, on energy resolution and on spatial resolution. The design energy resolutions of the Technical Design Report of BABAR were achieved.
  • Keywords
    Electrons , BaBar , pions
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1999
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number