Title of article
A comparison of time-of-flight and Larmor modulation neutron powder diffraction at a continuous reactor source
Author/Authors
Mulder، نويسنده , , F.M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
7
From page
296
To page
302
Abstract
A neutron powder diffraction instrument for application at a continuous reactor source based on time-of-flight (TOF) is compared with an instrument based on Larmor modulation (LM). For instruments with a similar wave-vector resolution (δq/q=4×10−3) it is found that the signal-to-noise ratios in a TOF experiment are generally better when using a wide wavelength neutron spectrum. This is caused by the fact that the noise in a LM experiment comes from the integrated neutron flux scattered in the detector, while in TOF the noise stems only from the intensity scattered into individual time channels.
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
1999
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2180309