Title of article :
High-pressure plastic scintillation detector for measuring radiogenic gases in flow systems
Author/Authors :
Schell، نويسنده , , W.R. and Vives-Batlle، نويسنده , , J. and Yoon، نويسنده , , S.R. and Tobin، نويسنده , , M.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
Radioactive gases are emitted into the atmosphere from nuclear electric power and nuclear fuel reprocessing plants, from hospitals discarding xenon used in diagnostic medicine, as well as from nuclear weapons tests. A high-pressure plastic scintillation detector was constructed to measure atmospheric levels of such radioactive gases by detecting the beta and internal conversion (IC) electron decays. Operational tests and calibrations were made that permit integration of the flow detectors into a portable Gas Analysis, Separation and Purification system (GASP). The equipment developed can be used for measuring fission gases released from nuclear reactor sources and/or as part of monitoring equipment for enforcing the Comprehensive Test Ban Treaty. The detector is being used routinely for in-line gas separation efficiency measurements, at the elevated operational pressures used for the high-pressure swing analysis system (2070 kPa) and at flow rates of 5–15 l/min [1, 2]. This paper presents the design features, operational methods, calibration, and detector applications.
Keywords :
Fission gases , Plastic scintillators , CTBT equipment , Monitoring , Calibration
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A